2.6 DEEP FREEZE:
DETAILS:
Name of the
Equipment
|
Deep Freeze DDF 1
|
Name of the Company
|
Blue Star
|
Model No.
|
HF 425 Ltr.
|
Serial no
|
42398
|
Year of Purchase
|
28/1/08
|
Date of First use
|
28/1/08
|
Inventory card no.
|
BJMC/Micro/Viro/NVBDCP/Deep
freeze/2008/18
|
Approximate Price
|
Rs.24600/-
|
AMC
|
Ramdev agency
|
Location
|
Dengue lab
|
Defrosting and cleaning
|
Once a month
|
Contents
|
Serum samples of virology laboratory
|
Responsibility
|
Laboratory technicians
|
Record maintained
in
|
DR- 05, DF-12, DF-16, PF-13
|
PURPOSE - Deep freeze is used to store serum sample for preservation
at -20ÂșC.
LOCATION – Dengue lab
DEFINITION - °C-
Degree centigrade
OPERATION:
- Put the Deep freezer in ‘ON’ condition.
- Adjust the required temperature i.e: -20°C.
- Monitor that the set temperature is maintained regularly.
- After the required temperature is attained, put the necessary material inside the deep freezer.
- Every morning note the temperature immediately by the first person opening it and record in the sheet pasted on the deep freeze and in register no PR 43.
MAINTENANCE:
- Record the freezer temp in the morning by the first person opening it.
- Temp chart should be maintained.
- Any technical problem should immediately be informed to the person in charge.
- They are connected with voltage stabilizer to protect the deep freezer from voltage fluctuations.
- Any operating problems like high temperature are recorded and brought to notice of officer concerned
PROCEDURE FOR DEFROSTING
- Every month, deep freeze should be defrosted, clean it with 1% Sod. Hypochlorite solution and dried with cloth. While defrosting & cleaning deep freeze no DDF-1, contents are transferred in deep freeze no 17 of polio lab.
RESPONSIBILITY – Laboratory technicians
SAFETY PRECAUTIONS-
- Do not modify or change system components.
- Disconnect unit from main power before attempting any maintenance to equipment or control.
REVIEW, RECORDS, REMARKS&RECOMMENDATIONS-
- PF 13 Temperature Record sheet file
- DR05 Log Book of Deep freeze.
- DF-12 Dengue AMC/Service/Repair Record file
- DF 16 Calibration file
REFERENCE:
- Manual of Laboratory Techniques (NICD), –chapter no 26.2.9, page no.212