2.5 ELISA READER
DETAILS:
Name of the
Equipment
|
ELISA reader
|
Name of the Company
|
Stat Fax ARK Company
|
Model No.
|
ARK 3200
|
Serial no
|
2416
|
Year of Purchase
|
10/04/2008
|
Date of First use
|
10/4//2008
|
Inventory card no.
|
BJMC/Micro/Viro/NVBDCP/ELISA reader/2008/31
|
Approximate Price
|
Rs.1,84,080/-
|
AMC
|
Ark Diagnostics
|
Location
|
Dengue lab
|
Responsibility
|
Laboratory technicians
|
Record maintained
in
|
DR- 03, DF-16, DF-12
|
PURPOSE - ELISA Reader is used to read and calculate the absorbance
values or optical density of the ELISA test.
LOCATION – Dengue Lab
DEFINITION: OD= optical density
OPERATION:
- Put the power switch on.
- Stat Fax 3200 will appear on screen.
- Select mode will appear on screen. In this window, Test option appears.
- Press test.
- Screen shows- Select stored test Enter test number.
- Press1.( 1st Filter 450 and 2nd filter 620nm wavelength)
- On screen following appears:
- Absorbance mode 8
- A to H format
- Load plate- read
- Put the plate on the reader platform.
- Press read.
- After the reading is over, take out the plate & put the switch off.
- If there is any spillage clean with 70% Ethanol, wipe with sterile cotton.
MAINTENANCE:
·
Suitable power supply is used and instrument is
placed with good air circulation around it, excessive vibration is avoided.
·
While using the instrument regularly attention
is paid to turn the power off and unplug the instrument.
·
Disposable gloves are used while handling the
instrument.
·
Outer surface of the instrument, the track and
the plate carrier are cleaned with a cloth dampened with water or mild
detergent.
·
Do not try to open the instrument.
·
If any problem is encountered, in charge is
informed immediately and company expert is called to solve the problem.
·
If there is any spillage of biohazard material,
turn the power off and unplug the instrument. By wearing disposable gloves
clean the outer surface of the instrument, the track and the plate carrier with
disposable cloth dampened either with 70% Ethanol.
·
If the biohazard material gets inside, call the
company expert for cleaning of the instrument.
RESPONSIBILITY – Laboratory technicians
SAFETY PRECAUTIONS -
- Operate the instrument gently.
- Do not use the instrument without a basic knowledge of its operation.
- For any kind of electrical problem, inform the section in charge & electrician immediately.
- Do not modify or change any program.
REVIEW, RECORDS, REMARKS & RECOMMENDATIONS
- DR 04 Log Book of ELISA reader
- DF 16 Calibration file
- DF-12
Dengue AMC/Service/Repair Record file
REFERENCE:
Manual of ELISA READER – (ARK ) Year-2008